Process Validation: Requirements and Implementation Conditions

Process Validation: Requirements and Implementation Conditions

Regulatory Requirements

ISO 13485 Section 7.5, “Production and Service Provision,” specifies requirements for process validation in Section 7.5.6. The standard states:

ISO 13485 7.5.6:

When the output from a process cannot be subsequently verified by monitoring or measurement, and defects therefore only become apparent after the product has been used or the service has been provided, the organization must perform process validation for the relevant manufacturing and service provision processes.

Similarly, FDA Quality System Regulation (QSR) §820.75 on Process Validation requires:

FDA QSR §820.75:

When the results of a process cannot be fully verified by subsequent inspection and test, the process must be validated with a high degree of assurance and approved according to established procedures.

Additionally, the EU Medical Devices Regulation (MDR) Article 15(3) contains comparable requirements, demonstrating international recognition of the necessity for validating manufacturing processes that cannot be fully verified through post-manufacturing inspection and testing. These regulatory requirements collectively clarify the fundamental conditions under which process validation becomes mandatory.

Core Conditions Requiring Process Validation

In essence, process validation is mandatory when post-manufacturing inspection and testing cannot adequately verify the output. Such processes are characterized by the involvement of destructive testing or inspection methods and are commonly referred to as “Special Processes” or “Critical Processes.”

Representative examples of special processes include: soldering, crimping, sterilization, adhesive bonding, welding, pressing, potting, implantation, coating, and surface treatments.

The defining characteristic of these processes is that the quality of the completed product cannot be fully verified through post-production inspection. Consider sterilization as a primary example. To confirm that sterilization has been properly executed, one must destroy the sterile barrier system (such as sterilization packaging) and test the internal biological indicators. However, once destructive testing is performed, that specific product can no longer be distributed as a medical device.

This fundamental dilemma necessitates sampling inspection for special processes involving destructive testing. Complete lot inspection is not feasible. For instance, while biological indicators can be evaluated by destructively testing a portion of products from a manufacturing lot, it is impossible to perform destructive testing on every single unit in the lot.

The Inherent Limitations of Sampling Inspection

The essential limitation of sampling inspection is that defects may exist in units that were not tested. Under no circumstances can incompletely sterilized or otherwise defective medical devices be released to the market. Patient safety and health are directly at stake.

Consequently, special processes cannot rely on post-manufacturing inspection. Instead, high-level quality assurance must be implemented prospectively to ensure that the process itself reliably produces acceptable results. This means that if 10,000 medical devices are manufactured, all 10,000 units must be completely sterilized without exception.

The fundamental purpose of process validation is to scientifically establish the adequacy of special processes prior to market distribution, thereby compensating for the inherent limitations of post-production inspection and ensuring patient safety.

Differential Inspection Capabilities in Medical Devices

Many medical devices can be inspected using non-destructive testing methods. Examples include visual inspection, oscilloscope measurement of electrical parameters, functional testing with automated testers, and power meter verification of output.

For such processes, post-manufacturing inspection provides adequate quality assurance, making prospective process validation less critical. Instead, in-process monitoring and post-manufacturing inspection—referred to as “process verification”—suffice to ensure quality.

Conversely, for special processes where adequate post-manufacturing verification is impossible, process validation must be performed beforehand. This represents the essence of a risk-based regulatory approach.

Practical Case Study

During an audit of a medical device manufacturing facility, the author observed an employee performing hand soldering with an incorrect soldering iron grip technique. Upon review of the employee’s training records, the soldering test score was found to be 60 out of 100 points.

In the audit report, the author flagged that personnel with a 60-point score should not perform soldering operations.

However, the audited department responded that “whether 60 points constitutes passing or failing is a matter of subjective interpretation, and the regulatory requirements contain no specific pass/fail criteria.”

This represents a fundamental misunderstanding. Soldering is a special process. Subsequent inspection methods—such as visual inspection or X-ray examination—cannot fully verify internal defects (voids, cold solder joints, burnout, etc.). Therefore, personnel performing soldering must possess a skill level that reliably ensures adequate results as a prerequisite of process validation.

An employee with near-perfect technical expertise would be expected to execute nearly all soldering connections correctly across 100 solder joints. Conversely, an employee scoring 60 points presents a statistical likelihood of 40% defect rate—meaning approximately 40 defective solder joints among 100 connections. This level of risk is unacceptable and poses extreme danger to patients.

The absence of specific criteria in regulatory requirements does not permit companies to set arbitrary standards based on subjective judgment. Rather, regulations require the establishment of technical competency levels based on scientific evidence, ensuring both process stability and product safety. This represents the practical significance of “process validation” as demanded by regulatory authorities.

Decision Matrix for Process Validation Necessity

The following matrix provides guidance for determining whether process validation is required:

Process CharacteristicPost-Manufacturing VerificationValidation RequirementRepresentative Examples
Destructive Testing InherentPost-manufacturing verification impossibleRequiredSterilization, welding, adhesive bonding, crimping
Non-Destructive Testing PossiblePost-manufacturing verification adequateNot requiredVisual inspection, dimensional measurement, electrical testing
Conditional VerificationRequires risk-based judgmentEvaluation necessaryHand soldering (partial verification possible, but complete verification impossible), potting, encapsulation

For processes with “conditional verification capability,” determining whether validation is necessary requires comprehensive evaluation of several factors: product criticality and potential patient harm, inspection method sensitivity (probability of detecting defects), specificity of inspectable parameters, inherent variation in manufacturing environment and personnel skills, and historical quality data. Scientific evaluation of these factors is essential to determine whether “process validation is necessary” or “process verification is adequate”—a critical responsibility for medical device manufacturers in regulatory compliance.

Emerging Regulatory Trends and Industry Implications

In recent years, the proliferation of AI/ML-based medical devices and digital health technologies has expanded the scope of special processes. For example, software integration, parameter configuration, and machine learning model validation are increasingly recognized as new forms of “special processes.” Concurrently, relevance to international standards such as IEC 62304 (Medical device software lifecycle processes) and IEC 62366-1 (Usability of medical devices) has deepened.

Medical device manufacturers must expand their process validation framework from traditional manufacturing-stage processes to encompass a broader perspective including design stages, software development phases, and cybersecurity components. The regulatory landscape continues to evolve, and corporate compliance strategies must advance correspondingly.

Fact Check Summary

Verified Accurate:

  • ISO 13485 7.5.6 requirements
  • FDA QSR §820.75 requirements
  • Fundamental concept of special processes requiring process validation
  • Relationship between destructive testing and sampling inspection limitations
  • Practical implications for process capability requirements

Enhanced Content:

  • Added EU MDR Article 15(3) reference for international regulatory balance
  • Integrated IEC 62304 and emerging AI/ML medical device regulatory considerations
  • Clarified “special process” definition for novice readers while maintaining technical precision
  • Expanded on risk-based decision-making framework
  • Added forward-looking industry trends and future regulatory implications

Maintained:

  • Original author’s intent, arguments, and professional perspective
  • Real-world practical case study with actual audit experience
  • Educational value for both beginners and specialists

Related post

Comment

There are no comment yet.